Background and Aim: Along with variety of implant systems, some manufacturers produced similar components for the main implant systems and
claimed them to be completely compatible with the main systems which
require more studies in this regard. Objective the aim of present study was to compare screw loosening in ITI and its compatible systems.
Materials and Methods:In this in vitro trial, 10 implants of ITI original and 10 abutments of the Cowell
and 10 Euroteknika systems were connected to ITI implants. abutment screw was tightened by 35 N/CM torque. After 5 minutes, the initial RTV of the specimens were measured. The connections were subjected to the cyclic loading (load of 75 N and 1 Hz) for 500000 cycles. Postcyclic RTV were measured by the torque wrench and the data were evaluated by ANOVA analyzing test.
Results:In ITI abutments; RTV was 30.7±2.26 and 23/0 N/CM before and after
cyclic loading. In Euroteknika abutments, RTV was 29.1±2.6 and 21/0
N/CM before and after cyclic loading. In Cowell system; RTV was decreased
from 29.2±3.61 to 21.641/2± N/CM before and after cyclic loading. No
significant differences were found regarding post-cyclic RTV in different
abutment/implant connections.
Conclusion:It seems , Under Similar Condition, the abutment brands of ITI, Euroteknika and
Cowell connected to ITI implant system did not significantly affected screw
loosening values following cyclic loading.
|