[Home ] [Archive]   [ فارسی ]  
:: Main :: About :: Current Issue :: Archive :: Search :: Submit :: Contact ::
Main Menu
Home::
Journal Information::
Articles archive::
For Authors::
For Reviewers::
Registration::
Contact us::
Site Facilities::
Indexing & Abstracting::
::
(CC-BY 4.0)

AWT IMAGE

Journal of Research in Dental Sciences is licensed under a "Creative Commons Attribution 4.0 International (CC-BY 4.0)"

..
Search in website

Advanced Search
..
Receive site information
Enter your Email in the following box to receive the site news and information.
..
:: Volume 14, Issue 1 (jrds 2017) ::
J Res Dent Sci 2017, 14(1): 1-6 Back to browse issues page
Comparison of the screw loosening in three implants systems
F Farid , H Haj mir agha , I Poornasiri *
, pournasiri_elnaz@yahoo.com
Abstract:   (6983 Views)

Background and Aim:  Along with variety of implant systems,  some manufacturers produced similar components for the main implant systems and

claimed them to be completely compatible with the main systems which

require more studies in this regard. Objective the aim of present study was to compare screw loosening in ITI and its compatible systems.

Materials and Methods:In this in vitro trial, 10 implants of  ITI original and 10 abutments of the Cowell

and 10 Euroteknika systems were connected to ITI implants. abutment screw was tightened by 35 N/CM torque. After 5 minutes, the initial RTV of the specimens were measured. The connections were subjected to the cyclic loading (load of 75 N and 1 Hz) for  500000 cycles. Postcyclic  RTV were measured by the torque wrench and the data were evaluated by ANOVA analyzing test.

Results:In ITI abutments; RTV was 30.7±2.26 and 23/0 N/CM before and after

cyclic loading. In Euroteknika abutments, RTV was 29.1±2.6 and 21/0

N/CM before and after cyclic loading. In Cowell system; RTV was decreased

from 29.2±3.61 to 21.641/2± N/CM before and after cyclic loading. No

significant differences were found regarding post-cyclic RTV in different

abutment/implant connections.

Conclusion:It seems , Under Similar Condition, the abutment brands of ITI, Euroteknika and

Cowell connected to ITI implant system did not significantly affected screw

loosening values following cyclic loading.

Keywords: Dental implant, torque, dental implant loading
Full-Text [PDF 12 kb]   (2017 Downloads)    
Type of Study: original article | Subject: Prothesis
Send email to the article author

Add your comments about this article
Your username or Email:

CAPTCHA


XML   Persian Abstract   Print


Download citation:
BibTeX | RIS | EndNote | Medlars | ProCite | Reference Manager | RefWorks
Send citation to:

Farid F, Haj mir agha H, Poornasiri I. Comparison of the screw loosening in three implants systems. J Res Dent Sci 2017; 14 (1) :1-6
URL: http://jrds.ir/article-1-761-en.html


Rights and permissions
Creative Commons License This work is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License.
Volume 14, Issue 1 (jrds 2017) Back to browse issues page
مجله تحقیق در علوم دندانپزشکی Res Dent Sci
Persian site map - English site map - Created in 0.05 seconds with 37 queries by YEKTAWEB 4645